Updated version of European breast tomosynthesis QC protocol (version 1.03)

Changes compared to version 1.01:

  • The dose reference values have been changed to limiting values.
  • The limiting dose value for 20 mm PMMA thickness has changed from 1.0 mGy to 1.2 mGy. This has been done because feedback was received and was noted that the value of 1.0 mGy was difficult to achieve with sufficient image quality for DBT systems.
  • Some text has been added regarding the mode in which mages have to be made in some methods of measurement.

 

Download the updated version here (bottom of the page).

Latest news

New Type Test Result

The HOLOGIC 3Dimensions has successfully passed the EUREF Type Test for 2D & Digital Breast Tomosynthesis mammography systems.

New Type Test Protocol is available

EUREF has developed a new Type Test Protocol (version 1.4) for both 2D & Digital Breast tomosynthesis mammography systems.

This protocol can be downloaded here.